Guo Zhi-rong, Yan Zong-yi. The Eccentric Effect of a Coaxial Viscometer[J]. Applied Mathematics and Mechanics, 1995, 16(4): 337-344.
Citation: Guo Zhi-rong, Yan Zong-yi. The Eccentric Effect of a Coaxial Viscometer[J]. Applied Mathematics and Mechanics, 1995, 16(4): 337-344.

The Eccentric Effect of a Coaxial Viscometer

  • Received Date: 1994-07-18
  • Publish Date: 1995-04-15
  • This paper studies the effect of the eccentricity between the inner bob and theouter cup in a coaxial viscometer on the measurements of the viscosity by using aperturbation scheme, Our numerical examples demonstrate that the error incurred bythe eccentricity would become larger as the characteristic Reynolds number decreasesor as the gap width between the bob and the cup gets narrower. However, such errorsmay not be signifcant if the eccentric distance is not larger than 1/10 of the gapwidth for the usual working range of a coaxial viscometer.
  • loading
  • [1]
    严宗毅、魏茂元、于天文,《血液流变学—基础、检测、应用》,黑龙江科学技术出版社(1993), 13.
    [2]
    吴望一,《流体力学》.下册,北京大学出版社(1983), 213,
    [3]
    李平、严宗毅、庄逢源,双筒粘度计中牛顿流体的振荡解,应用数学和力学,14(6) (1993),535-541.
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views (1821) PDF downloads(451) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return